How to
Identify a Non-Magnetic Gas Cylinder?
The safest approach is not to identify a cylinder by
appearance, color, or a handheld magnet alone. Identification should be
based primarily on the cylinder's manufacturer markings, material
specification, documentation, and MRI/MR safety labeling.
1. Check the cylinder material marking
Look at the permanent markings stamped or engraved around
the cylinder shoulder or neck. A non-magnetic cylinder may be manufactured from
materials such as:
- Aluminum
alloy commonly used for cylinders intended for MRI environments.
- Certain
austenitic stainless steels — these may have low magnetic response,
although some stainless steels can become partially magnetic after
manufacturing or cold working.
- Other
specialized non-ferromagnetic alloys approved by the manufacturer.
The cylinder should have permanent identification showing
information such as manufacturer, specification/standard, serial number,
working pressure, test date, and material designation.
2. Look for MRI/MR safety identification
For cylinders intended to enter an MRI environment, look for
specific labeling indicating that the complete cylinder assembly is
suitable for that environment. Depending on the equipment and applicable
standard, it may be identified as MR Safe or MR Conditional.
This distinction is important:
- MR
Safe means the item poses no known hazards in all MR environments
under the applicable definition.
- MR
Conditional means it is safe only under specified MRI conditions.
- An
item without appropriate MR safety information should not automatically
be considered safe for an MRI room.
3. Do not check only the cylinder body
Even when the cylinder body is aluminum or another
non-ferromagnetic material, other components can contain ferromagnetic
materials. Check the complete assembly, including:
Cylinder body → valve → regulator → pressure gauge → flowmeter
→ protective guard/cap → trolley or cylinder stand
For example, an aluminum oxygen cylinder fitted with a
ferromagnetic regulator or placed on a conventional steel trolley can still
present a serious MRI projectile hazard.
4. Magnetic testing is only a supplementary check
A controlled magnetic-response test may be used as part of
an organization's approved MRI screening procedure. However, simply placing a
household magnet against the cylinder is not sufficient certification
that the cylinder is MRI-safe.
A cylinder that shows little attraction at one point may
still contain magnetic components elsewhere. More importantly, MRI suitability
depends on the entire assembly and the conditions under which it was tested,
not just whether the cylinder body attracts a small magnet.
Practical Identification Procedure for Hospital Staff
Before allowing a medical gas cylinder into an
MRI-controlled area, use a procedure such as:
Step 1: Read the cylinder's permanent manufacturer
and material markings.
Step 2: Confirm the cylinder specification and material against
manufacturer documentation.
Step 3: Verify its MR Safe/MR Conditional status and any
conditions of use.
Step 4: Inspect the valve, regulator, gauge, flowmeter, guard and
trolley separately.
Step 5: Confirm that the complete assembly is listed in the hospital's
approved MRI equipment inventory.
Step 6: Apply the facility's approved identification/tagging system
after verification.
Step 7: If identification or documentation is uncertain, do not take
the cylinder into the MRI magnet room (Zone IV) until it has been cleared
by the MRI safety responsible person.
Important Point for Oxygen Cylinders
A cylinder being aluminum does not automatically mean
that the complete oxygen-cylinder assembly is MRI-safe. Similarly, cylinder
color identifies the gas according to applicable local standards; it does not
establish magnetic compatibility.
For hospital Facilities/MEP and Medical Gas teams, a good
identification tag would include:
NON-FERROMAGNETIC / MRI USE
Gas: Medical Oxygen
Cylinder Material: Aluminum Alloy
MR Status: MR Safe / MR Conditional
Regulator/Valve: Verified
Asset/Serial No.: ______
Inspection Date: ______
Verified By: ______
The key rule is verifying the complete cylinder assembly from its markings and manufacturer MR-safety documentation rather than relying on cylinder color or a simple magnet test.
Important: Never take an unverified gas cylinder into
an MRI room (Zone IV).